beautypg.com

Warning, Caution – Brewer LiftMate User Manual

Page 5

background image

5

Document # 2102196 RevA

WARNING

Please observe the following

warnings (continued).

● Patient must stand in the center of the platform during the lift

procedure. Never lean forward, backward or sideways beyond

the area indicated by the dashed line.

● Do not allow more than one person on the LiftMate at a time.

● Ensure that the back of the LiftMate is pushed as close to the

exam table as possible.

● Make sure that all exam table drawers, stirrups, attachments

or other protrusions are pushed in to prevent damage.

Children, Caregivers and Untrained Personnel:

● Anyone not trained in the safe use of the LiftMate must not be

allowed to operate it.

● Ensure that children and others are clear of the LiftMate at all

times whether it is being operated or not.

● Never allow the patient to operate the LiftMate at any time.

Special Patient Situations:

● The LiftMate must not be used by patients to pull themselves

up out of a wheelchair or walker.

● Do not use the LiftMate for patients that are unable to support

their own weight.

● Never exceed LiftMate’s maximum capacity of 500 lbs.

General: Please observe the following consideration.

● LiftMate must not be used to transport anything including

patients.

● If battery fails while casters are up or down, plug into receptacle

to recharge.

WARNING

Please observe the following

warnings.

Patient and Operator: Ensure that loose clothing, buttons,

jewelry or other such items that can snag onto the LiftMate

are kept clear of catch points
Patient:

● Never put a patient on the LiftMate while casters are

extended.

● Patient should only get on or off of the LiftMate when

all four base feet are firmly on the ground.

● Do not allow patients to pull themselves up by using

any part of the LiftMate.

WARNING

Never place any body parts between

Danger

Zone

(Cautions on next page)

Top View

Danger Zone

CAUTION

Do not use any power supply other than that

listed on the rating label. Failure to do so

may result in serious injury and/or equipment

damage.

This product has been evaluated with respect to

electrical shock, fire, and mechanical hazards

only in accordance with UL60601-1; IEC60601-1;

CAN/CSA C22.2 No. 601.1

ELECTRICAL REQUIREMENTS

Electromagnetic Interference

This product is designed and built to minimize electromag-

netic interference with other devices; however, if interfer-

ence is noticed between another device and this product,

remove the interfering device from the room or plug this

product into an isolated circuit.

Do not perform any procedures while patient is on

the LiftMate.

Use 115 VAC, 60 Hz as noted on the rating label.

Failure to do so may result in electrical shock to

personnel and will result in damage to the LiftMate.

Do not use the LiftMate in an explosive or oxygen

enriched atmosphere. Failure to do so may result in

serious personal injury or death.

Charging the battery with any device other than

through the Control Box on the unit may result in

serious personal injury, damage to other LiftMate

components, and/or premature failure of the battery.

WARNING

the Handrail Crossbrace and Shrouds, or under the

front edge of the Platform.

Crossbrace

Crossbrace

Both views show platform raised

Shroud

Platform

Edge