beautypg.com

Emergency stops – Brewer LiftMate User Manual

Page 10

background image

10

Document # 2102196 RevA

UP

DOWN

LOW BATTERY

OPERATION (continued)

Patient Handling with the LiftMate

(Lowering the Patient)

1. Roll the back of the LiftMate against foot end or side

of exam table, depending on position of patient on

the table.

2. Ensure patient’s feet are clear of the platform and

any portion of the exam table as it is raised.

3. Make sure the platform is raised by pushing the “UP”

button, Figure 8.

4. Have patient slide forward to stabilize feet onto

platform.

5. Patient can then come to a standing position on

platform and grasp handrail.

6. Lower platform all the way down, Figure 9, and assist

patient off of LiftMate.

Emergency Stops

Emergency Stops are located on the outer right-hand and outer left-hand Shrouds, Figure 10. In the unlikely event of

unintended movement of the Platform, depressing the Emergency Stop, Figure 11, will cause all Platform movement

to cease. Have a qualified technician inspect the unit prior to using. Rotating the Emergency Stop Switch clockwise,

Figure 12, will reset it to normal operation. Note: The Emergency Stops only deactivate the Platform. The casters will

still deploy with the Emergency Stop button depressed.

Figure 10. Emergency

Stop Locations

Figure 11. Activating the

Emergency Stop

Figure 12. Resetting the

Emergency Stop

Figure 9.

Emergency Stops