Adjustment procedure, Adjustment procedure 5-23 – Agilent Technologies 83438A User Manual
Page 77

5-23
Servicing
Adjustment Procedure
Adjustment Procedure
The Agilent 83438A has only one adjustment procedure, total power. Periodic
adjustment is not required to maintain safety.
C A U T I O N
Option 022 instruments have an angled-fiber output. Be sure to use an angled-
fiber patchcord during testing. If you do not have an angled-fiber patchcord
available, you can purchase the required accessories from Agilent
Technologies. These include the Agilent 81000SI DIN connector interface and
the Agilent 81113PC DIN 4108 (angled) to Super-PC patchcord.
W A R N I N G
To avoid exposure to the laser path of a CLASS IIIb LASER
PRODUCT, do not open the A2A1 assembly. There are no
serviceable components inside. Do not damage the fiber-optic
cable that connects the CLASS IIIb pump laser, A2A3, to the
A2A1 assembly. Other laser paths outside of the A2A1
assembly do not require precautions to maintain safety.