MAINTENANCE
Routine Inspection: Not required except for periodic inspection of the cable and connector to
ensure that these are neither damaged nor softened by incompatible liquid.
TABLE 1
ELECTRICAL CONNECTIONS
4-20mA
VOLTAGE
TYPE
NUMBER
CE
CONNECTOR
IN+
IN-
EARTH
IN+
COM
OUT+
EARTH
512X-E1
Yes
4 Pin Mini DIN
1
2
E
1
2
3
E
512X-NA
Yes
IP67 Weatherproof
Cable Gland
R
BK
DRAIN
R
BK
W
DRAIN
512X-N1
Yes
IP65 Cable
R
BK
DRAIN
R
BK
W
DRAIN
516X-B1
Yes
8-4 Bayonet
Connector
A
B
D
A
C
B
D
516X-B3
Yes
10-6 Bayonet
Connector
A
B
E
A
C
B
E
516X-E2
Yes
Large DIN 436
Connector with
Mating Connector
1
2
E
1
2
3
E
516X-E2
Yes
½” Connector with 1
meter lead
R
BK
DRAIN
R
BK
W
DRAIN
OUTPUT
SUPPLY VOLTAGE
11 4 to 20mA
7 to 35V d.c
28 1 to 6V
7.5 to 35V d.c
2R 1 to 11V
12.5 to 35V d.c
27 1 to 5V
6.5 to 35V d.c
24 0.5 to 5.5V
7 to 35V d.c.
2B 0 to 5V
6.5 to 35V d.c.
G 0 to 10V
11.5 to 35V d.c.