4.3 current output Zero adjustment

While monitoring the current output, and with both pressure ports open to atmosphere, the zero may

be adjusted. For unidirectional pressure ranges, turn the zero adjustment screw until a reading of 4 mA

(±.08 mA) is achieved. For bidirectional ranges, set zero to 12 mA (±.08 mA).

4.4 current output span adjustment

Span or full scale output adjustments should only be performed by using an accurate pressure standard

(electronic manometer, digital pressure gage, etc.), with at leads comparable accuracy to the 230 series

(±.25% full scale). With full range pressure applied to the high pressure port (reference port open to

atmosphere), adjust span to achieve 20 mA (±0.08 ma) output.

Example 1:

Unidirectional pressure range of 0 to 1 PSI

Apply 1 PSI, adjust output to 20 mA

Example 2:

Bidirectional pressure range of ±.5 PSI

Apply .5 PSI, adjust output to 20 mA

5.0 RETURNING PRODUCTS FOR REPAIR

Please contact a Setra Systems application engineer (1-800-257-3872, 1-978-263-1400) before returning unit

for repair to review information relative to your application. Many times only minor field adjustments may

be necessary.

When returning a product to Setra Systems, the material should be carefully packaged and accompanied by

our service return form found at http://www.setra.com/uploads/service return form.pdf., completely filled

out, and shipped prepaid to:
Setra

Systems,

Inc.

159

Swanson

Road

Boxborough,

MA

01719-1304

Attn:

Repair

Department

To assure prompt handling, please supply the following information and include Service Return Form inside

the packaged returned materia:

1. Name and phone number of person to contact.

2. Shipping and billing instructions.

3. Full description of malfunction.

4. Identify any hazardous material used with product.

Notes: Please remove any pressure fittings and plumbing that you have installed and enclose any required

mating electrical connectors and wiring diagrams.

Allow approximately 3 weeks after receipt at Setra Systems for the repair and return of the unit. Non-warran-

ty repairs will not be made without customer approval and purchase order to cover repair charges.

Calibration Services

Setra maintains a complete calibration facility that is traceable to the National Institute of Standards &

Technology (NIST). If you would like to recalibrate or recertify your Setra pressure transducers, please call our

Repair Department at 800-257-3872 (978-263-1400) for scheduling.