ZERO OUTPUT
Output can be adjusted to zero by potentiometer as shown in diagram. Unit factory adjusted to
0.00 VDC output (
±
5mV).
SPAN
Can be adjusted by potentiometer as shown in diagram. Unit factory adjusted to 5.00 VDC full
scale output (
±
5mV).
LINEARITY - DO NOT TOUCH
Factory adjusted for best linearity.
Circuit Board 1
Circuit Board 2
NOTE: When circuit cover is removed, a zero shift can be noticed. The zero will return by the same
amount when the cover is replaced. This displacement of the output should be taken into account when
making adjustments or during recalibration
*Adjustments: Due to the high accuracy of this transducer, adjustments to zero and span should only be
made with the use of primary pressure standards of known accuracy higher than that of the Model 270.
- Model ASM (8 pages)
- Model 204 (4 pages)
- Model 209 (4 pages)
- Model 227 (8 pages)
- 264 (4 pages)
- 269 (4 pages)
- Model 280E (4 pages)
- Model 370 (48 pages)
- Model 526 (7 pages)
- Model 540 (9 pages)
- Model 595 (8 pages)
- Model 869 (28 pages)
- Model 321 (4 pages)
- Model SRH (12 pages)
- Model Datum 2000 (31 pages)
- Model 217 (8 pages)
- Model 230 (12 pages)
- Model 242901-06 (1 page)
- Model 265 (4 pages)
- Model 299 (4 pages)
- Model 470 (35 pages)
- Model 524 (6 pages)
- Model 546 (8 pages)
- Model 730 (16 pages)
- GCT-225 (8 pages)
- Model 141 (4 pages)
- Model 205-2 (4 pages)
- Model 223 (8 pages)
- Model 231 (12 pages)
- Model 256 (8 pages)
- Model 276 (2 pages)
- Model 3100 (8 pages)
- Model 516 (6 pages)
- Model 550 (9 pages)
- Model 760 (18 pages)
- Model LD330 (4 pages)
- Model SRMD (8 pages)
- Model 201 (4 pages)
- Setra Model 206 (4 pages)
- Model 224 (8 pages)
- Model 231RS (16 pages)
- 260 (2 pages)
- 267MR (14 pages)
- Model 278 (2 pages)