Setra System Model C280E User Manual
Page 3
Span Adjustment
Span can be adjusted approximately ±0.5 mA from the nominal 16 mA full scale span. The
span adjustment potentiometer is located as shown in the diagram, and may be adjusted
without affecting linearity calibration.
Other Adjustment
Adjusted at factory. Do not touch.
RETURNING PRODUCTS FOR REPAIR
Please contact Setra (1-800-257-3872, 978-263-1400) before returning unit for repair to review
information relative to your application. Many times, only minor field adjustments may be neces-
sary.When returning a product to Setra, the material should be carefully packaged and shipped
prepaid to:
Setra Systems, Inc.
159 Swanson Road
Boxborough, MA 01719
Attn: Repair Department
To assure prompt handling, please supply the following information and include it inside the
package of returned material:
1. Name and phone number of person to contact.
2. Shipping and billing instructions.
3. Full description of the malfunction.
4. Identify any hazardous material used with product.
Notes:
Please remove any pressure fittings and plumbing that you have installed and enclose
any required
mating electrical connectors and wiring diagrams.
Allow approximately 3 weeks after receipt at Setra for the repair and return of the unit. Non-war-
ranty repairs will not be made without approval and a purchase order to cover repair charges.
Calibration Services
Setra maintains a complete calibration facility that is traceable to the National Institute of
Standards & Technology (NIST). If you would like to recalibrate or recertify your Setra pressure
transducers or transmitters, please call our Repair Department at 1-800-257-3872 (978-263-1400)
for scheduling, cost and turnaround estimates.
Thank you,
Setra Systems, Inc.