Tuchel connector pin configuration, Draft, E-m-hm4-v0_08 – ROTRONIC HM4 User Manual
Page 24: 3 tuchel connector pin configuration

E-M-HM4-V0_08
ROTRONIC AG
Bassersdorf, Switzerland
Document Code
Unit
HM4 Humidity and Temperature Probe with
Heated Sensor
Instruction Manual
Document Type
Page 24 of 41
Document Title
© 2014; ROTRONIC AG
E-M-HM4-V0_08
6.3 Tuchel Connector Pin Configuration
The HM4 probe is available with Tuchel connector. It is connected either directly at the housing or remotely at
the end of a cable.
Figure 5: Tuchel connector and pin numbering
Pin
Name
Function
1
V (+)
5…24 VDC
15…24 VDC, depending on output variant and functions used
2
GND
Supply ground
3
AGND
Analog GND
4
OUT2, analog
Temperature (default setting: -
40…60 °C)
5
RS-485 +
RS-485 TX+ / RX+
6
OUT1, analog
Humidity (default setting: 0…100 %RH)
7
RS-485 -
RS-485 TX- / RX-
Guide nose
1
2
3
1
4
3
1
1
5
3
1
1
6
3
1
1
7
3
1
1
DRAFT