User-Specific Settings and Functions
The HM4 has a variety of functions and settings that can be adapted by users to their requirements. This chapter
contains an overview of all functions and their default settings.
4.1 General AirChip3000 Functions
This section outlines the general functions of the AirChip family available in the HM4. All functions can be used
with the HW4 software and are described in detail in the manual E-M-HW4v3A2-001.
4.1.1
Humidity and Temperature Measurement
Humidity and Temperature Adjustment
One-point or multi-point humidity calibration and adjustment
One-point and two-point temperature calibration and adjustment
Time stamp for every calibration and adjustment point
Saving and display of the last adjustment data and values
Creation of a calibration and adjustment log
IMPORTANT:
All heating and cleaning functions must be switched off during calibration and adjustment!
4.1.2
Device Validation
Simulation Mode
Defines freely selectable values at the analog and digital outputs. Simulation mode is enabled/disabled with
the HW4 software.