beautypg.com

Iv. specifications – Dwyer PMT User Manual

Page 5

background image

Particulate Monitoring Systems

Installation & Operating Manual


Document No. 209-1026-E

v

©2007

IV. Specifications

ELECTRICAL

PARAMETER DETAIL

SPECIFICATION NOTE

Measurement Units

picoamperes (pA)

1 x 10 ⎯ ¹² Amp

Standard

10.0pA – 5000pA

Measurement Range

Optional

5.0 and 0.5pA – 5000pA

Accuracy

± 5% of Range

Full Temp Range

Operating

-15F to +160F (-25C to +70C)

Ambient Temperature
Range

Storage

-40F to +185F (-40C to +85C)

Digital Filtering

Range

Ajustable 0.5 Sec – 60 Sec

Isolation

500VDC Process to Loop

Output
Compliance

Max. RLoad = (Vsupply–18V)/0.02A
(24VDC = 300Ohm Rload Max.)

Scale

Linear or Logarithmic

4-20mA Output

pA Range

0-100, 0-500, 0-1000, 0-5000 Linear
0.5–500, 5–5000 Logarithmic

MECHANICAL

Standard

½” NPT Mount

Mounting

Optional Tri-clamp,

Flange

Other, consult
factory

Sensor

Teflon over Stainless

Materials

Body Stainless

Steel

-40F to +250F (-40C to +120C)

-40F to +450F (-40C to +232C)

Process Temperature
Range

Operating

-40F to +800F (-40C to +427C)

two piece
configuration
only

Pressure Range

Operating

Full Vacuum to 10PSI

Other, consult
factory

Type

Painted Cast Aluminum

Enclosure

Rating

NEMA 4X (IP 66)

OPERATING and APPLICATION RANGE

10.0pA

5.0pA
Optional

At least 10.0-5000 mg/m³
At least 0.004 to 2.0 gr/cf
At least 5.0 – 5000 mg/m³
At least 0.002 to 2.0 gr/cf

Visible

Barely Visible
To Visible

Particle Concentration
(Approximate Guide)

0.5pA
Optional

At least 0.5 to 500 mg/m³
At least 0.0002 to 0.2 gr/cf

Barely Visible
To Invisible

Particle Velocity

300 ft/min. (91 m/min.) and Higher

Particle Size

0.3 Micron and Higher