beautypg.com

Dwyer IS626 User Manual

Dwyer instruments, inc, Series is626 pressure transducer

background image

The Dwyer Series IS626 Pressure Transducer convert pressure into a
standard 4-20 mA output signal. The Series IS626 can be used to accu-
rately measure compatible gases and liquids. Series IS626 full scale ac-
curacy is 0.25% (see specifications). Designed for industrial
environments with a NEMA 4X (IP66) housing, this transmitter resists
most effects of shock and vibration.

Intrinsic Safety Approval Classification
The IS626 is UL listed for use in Hazardous (Classified) Locations. The
protection method is by Intrinsic Safety, “ia”. It was investigated by UL
under UL Standard 913 Sixth Edition and CSA Standard No. 157-92.

For use in Hazardous (Classified) Locations:
Class I Div. 1 Groups A,B,C,D
Class II Div. 1 Groups E,F,G
Class III Div. 1
Temperature Code: T4 @ 80°C ambient
Install in accordance with control drawing 01-700797-00.

WARNING: To prevent ignition of flammable or
combustible atmospheres, disconnect power be-
fore servicing.
Use with approved safety barriers using entity eval-
uation.

Series IS626 Pressure Transducer

Specifications - Installation and Operating Instructions

Bulletin E-IS-626

SPECIFICATIONS
Service:
Compatible gases and liquids.
Wetted Materials: Type 316, 316L SS.
Accuracy: 0.25% full scale (includes linearity, hysteresis, and repeata-
bility).
Temperature Limit: 0 to 176°F (-18 to 80°C).
Compensated Temperature Range: 0 to 176°F (-18 to 80°C).
Thermal Effect: ±0.02% FS/°F (includes zero and span).
Pressure Limits: See Pressure Range Table (Next Page).
Power Requirements: 10 to 28 VDC.
Output Signal: 4 to 20 mA.
Response Time: 50 msec.
Loop Resistance: 0 - 900 ohms maximum.
Current Consumption: 38 mA (maximum).
Electrical Connections: 3 ft. cable or 4-pin M-12 Connector.
Process Connection: 1/4˝ male/female NPT and BSPT.
Enclosure Rating: NEMA 4X (IP66).
Mounting Orientation: Mount in any position.
Weight: 8.9 oz (252 g).
Agency Approvals: CE, UL Intrinsically Safe to UL Standard 913. (See
Intrinsic Safety Approval Classification)
The following standards were used for CE approval:
IEC 61000-4-2: 2001
IEC 61000-4-3: 2006
IEC 61000-4-4: 2004
IEC 61000-4-5: 2005
IEC 61000-4-6: 2006
IEC 61000-4-8: 2001
CENELEC EN 55011: 2003
CENELEC EN 61326: 2003
89/336/EEC EMC Directive

MALE NPT/BSPT CONNECTOR WITH CABLE GLAND

1/4 NPT or
1/4 BSPT

1-1/16 [26.99] HEX

5-3/16 [131.75]

6-1/4 [158.75] MAX

ø1 [25.40]

1/4 NPT OR

1/4 BSPT

1-1/16 [26.99] HEX

MALE NPT/BSPT CONNECTOR WITH MALE M-12 CONNECTOR

5-1/16 [128.58]

5-21/32 [143.67]

ш1 [25.40]

ш1 [25.40]

1-1/16 [26.99] HEX

5-1/16 [128.58]

5-21/32 [143.67]

FEMALE NPT/BSPT CONNECTOR WITH MALE M-12 CONNECTOR

1/4 NPT OR

1/4 BSPT

FEMALE

CD

= CRITICAL DIMENSION
STANDARD TOLERANCES UNLESS NOTED:
ALL DECIMAL DIMENSIONS ± /005
ALL ANGLES ± 1’

FEMALE NPT/BSPT CONNECTOR WITH CABLE GLAND

1-1/16 [26.00] HEX

5-3/16 [131.76]

5-1/4 [158.75]

1/4 NPT OR

1/4 BSPT

FEMALE

ø1 [25.40]

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

Vmax
28VDC

Imax
93mA

Ci
0.051_F

Li
240_H

IS626 Entity Parameters

Pi
0.651W

Project3:Layout 3 12/7/09 11:21 AM Page 1