Dwyer ISDP User Manual
Page 15
ZERO
and
SPAN
(Calibration of Zero and Span) SUB MENUS
The lower display reads
CAL
in this mode.
ZERO
Calibration
NOTE: For accurate calibration, DO NOT apply any pressure when performing
this function.
With the display reading
ZERO
, press the ENTER key. The upper display will blink.
Press ENTER again to complete the zeroing of the instrument or press the
MENU
key to cancel.
SPAN
Calibration
With the display set to
SPAN
, apply full scale pressure to the unit. Press the ENTER
key. The upper display will blink. Press ENTER again to complete the calibration or
press the
MENU
key to cancel.
Maintenance
Upon final installation of the Series ISDP intrinsically Safe Differential Pressure
Transmitter, no routine maintenance is required. The Series ISDP is not field serviceable
and should not be returned if repair is needed (field repair should not be attempted and
may void warranty). Be sure to include a brief description of the problem plus any rele-
vant application notes. Contact customer service to receive a return goods authorization
number before shipping.