As received / as left report, Reset to factory defaults – Crystal ConFigM30 Digital Pressure Calibrator Software User Manual
Page 11

Settings
9
ConfigM30 User's Manual
To remove an existing password, enter the current password and click the
Remove Password button
. This will clear the password and not require it to be
entered on future gauge updates.
If you have forgotten the password, it is possible to clear the password by requesting an unlock code from the factory. Contact
to obtain
the code. Once you have the code, click the
Forgot Password button
and follow the on-screen directions to remove the password.
—
AS RECEIVED / AS LEFT REPORT
ConfigM30 has a built-in report which details any changes you have made to the gauge. Clicking the As Received / As Left Report button will bring up a report
of any changes made to the gauge. From this dialog, you can print the report or save it to a HTML file which can be viewed or printed later using any web
browser, such as Microsoft Internet Explorer.
—
RESET TO FACTORY DEFAULTS
When ConfigM30 first makes changes to the gauge, it makes a copy of the settings in the gauge. It is possible to use those values to return the gauge to its
factory condition by simply pressing the
(
Load Factory Configuration
)
button. This will return the gauge to the condition as shipped from the factory.
!
WARNING: Returning the gauge to factory defaults will also change the userspans back to their original values. If you wish to preserve the userspan
values, you must write them down before restoring the gauge to factory defaults, and then re-enter the userspan values.