gpm

Average Frequency

**UUT Hz Counts/GAL

UUT K (Hz × 60)/NK gpm

Error (% FS)

15 00

769 7

3078 59

14 90

–0 65

9 06

466 1

3086 75

9 03

–0 38

5 49

285 2

3118 64

5 52

0 65

3 32

171 7

3103 95

3 32

0 17

2 00

103 6

3101 80

2 01

0 10

**UUT = Unit Under Test

If this meter produced an actual linear output, the K factor calculation for the Unit Under Test would be exactly the same for

each measurement point Inspection of the UUT K in the example provided shows that this is not the case and indicates that

this meter is not a perfectly linear device
Many programmable displays allow for linearization and can provide a better match of the displayed flow values with the

actual flow values by incorporating more measurement points In the example, the unit would be programmed for six points,

the five data points and a zero point, and use pairs of input values to accomplish the linearization

DEFAULT K FACTOR VALUES

Meter Size

Default K factor

Lower Limit

Upper Limit

3⁄8 in

20,000

16,000

24,000

1/2 in

13,000

10,400

15,600

3/4 in

2750

2200

3300

7⁄8 in

2686

2148

3223

1 in

870 0

696 0

1044

1-1/2 in

330 0

264 0

396 0

2 in

52 0

41 6

62 0

3 in

57 0

45 6

68 0

4 in

29 0

23 2

35 0

6 in

7 0

5 6

8 0

8 in

3 0

2 4

4 0

10 in

1 6

1 3

2 0

Programming Manual

Page 19

November 2013