Default k factor values, Default k factor values 19 – Badger Meter B2800 User Manual
Page 19

gpm
Average Frequency
**UUT Hz Counts/GAL
UUT K (Hz × 60)/NK gpm
Error (% FS)
15 00
769 7
3078 59
14 90
–0 65
9 06
466 1
3086 75
9 03
–0 38
5 49
285 2
3118 64
5 52
0 65
3 32
171 7
3103 95
3 32
0 17
2 00
103 6
3101 80
2 01
0 10
**UUT = Unit Under Test
If this meter produced an actual linear output, the K factor calculation for the Unit Under Test would be exactly the same for
each measurement point Inspection of the UUT K in the example provided shows that this is not the case and indicates that
this meter is not a perfectly linear device
Many programmable displays allow for linearization and can provide a better match of the displayed flow values with the
actual flow values by incorporating more measurement points In the example, the unit would be programmed for six points,
the five data points and a zero point, and use pairs of input values to accomplish the linearization
DEFAULT K FACTOR VALUES
Meter Size
Default K factor
Lower Limit
Upper Limit
3⁄8 in
20,000
16,000
24,000
1/2 in
13,000
10,400
15,600
3/4 in
2750
2200
3300
7⁄8 in
2686
2148
3223
1 in
870 0
696 0
1044
1-1/2 in
330 0
264 0
396 0
2 in
52 0
41 6
62 0
3 in
57 0
45 6
68 0
4 in
29 0
23 2
35 0
6 in
7 0
5 6
8 0
8 in
3 0
2 4
4 0
10 in
1 6
1 3
2 0