beautypg.com

Thermo Fisher Scientific Alpha pH 190 (In the process of updating the brand guideline & product image) User Manual

Page 32

background image

- 26 -

9.7 Viewing pH Electrode Data – P5.0

In pH mode, Program 5 has three “view only” options that lets you
check the electrode parameters for diagnostic purposes.
P5.1: view pH electrode offset
P5.2: view pH electrode slope
P5.3: view temperature probe offset (ATC on only)

These parameters will change each time you recalibrate the
controller.

P5.0

elec

Slpe

P5.1

P5.2

P5.3

p.ofs

t.ofs

0.0

5 .

92

5

.

0

Press ▲ and ▼ keys together (ESCAPE) at anytime, to leave the
Setup mode and return to Measurement mode.

P5.0: Viewing pH electrode data
1.

Key in the password “022” as per procedure in Section 5.

2. Press

▲ key four times. The screen will scroll P5.0 and ELEC.

Press ENTER.

3. The screen will scroll P5.1, then P.OFS, then pH electrode

offset in mV at the pH 6.86 / 7.00 calibration point.

4. Press ENTER. The screen will display P5.2, then SLPE, then

the electrode slope in mV.

5.

MTC on: press ENTER to return to general Setup mode.
ATC on: go to next step.

6. The screen will scroll P5.3, then T.OFS, then the temperature

offset value.

7.

Press ENTER to return to general Setup mode.
Press ▲ and ▼ keys together (ESCAPE) to return to
measurement mode.