P3.3: Select temperature compensation
This controller features selectable automatic temperature
compensation (ATC) or manual temperature compensation (MTC).
ATC automatically compensates for temperature fluctuations; MTC
let you select a specific value at which temperature will be
compensated. ATC readings require a temperature sensor.

1.

Key in the password “022” as per procedure in Section 5.

2. Press

▲ key twice. The screen will scroll P3.0 and CONF.

Press ENTER three times.

3.

The screen will scroll P3.3 and MTC (or ATC).

4. Press

▲ or ▼ keys to toggle between MTC and ATC.

5.

Press ENTER to confirm and return to general Setup mode.
Press ▲ and ▼ keys together (ESCAPE) to return to
measurement mode.

NOTE: ORP is not affected by temperature and so ATC does not
operate in ORP mode. When set for ATC, the pH 190 will display
temperature if a 100 Ω Pt 100 sensor is connected.