9.2 Setup mode overview

P1.0: Set Point 1

P1.1: select relay 1 set point value

P1.2: select relay 1 as low or high set point

P1.3: set relay 1 hysteresis value


P2.0: Set Point 2

P2.1: select relay 2 set point value

P2.2: select relay 2 as low or high set point

P2.3: set relay 2 hysteresis value


P3.0: Configuration

P3.1: select pH or ORP units

P3.2: select symmetrical or asymmetrical input mode

P3.3: select temperature compensation (MTC or ATC)


P4.0: Calibration data

P4.1: view first calibration point

P4.2: view second calibration point (pH mode only)


P5.0: Electrode data

P5.1: view electrode offset

P5.2: view electrode slope (pH mode only)

P5.3: view temperature probe offset (ATC mode only)


P6.0 Controller reset

P6.1: clear calibration data and revert to factory default settings