Thermo Fisher Scientific Alpha pH 190 (In the process of updating the brand guideline & product image) User Manual

Page 27

background image

- 21 -

9.4 Set Point 2 – P2.0

Setup program P2.0 allows you to set parameters for relay 2.

P2.1: select relay 2 set point value
P2.2: select relay 2 as low or high set point
P2.3: set relay 2 hysteresis value (dead band)

P

2.0

P

2.1

P

2.2

P

2.3

10.00

1.00

HYS 2

SP 2

SP 2

SP 2

LOW

HIGH


Press ▲ and ▼ keys together (ESCAPE) at anytime, to leave the
Setup mode and return to Measurement mode.

P2.1: Select relay set point value
This lets you choose the pH (or ORP) value that will cause your
controller to activate. If this value is crossed, the set point relay 2
LED will light.
1.

Key in the password “022” as per procedure in Section 5.

2. Press

▲ key once. The screen will scroll P2.0 and SP2. Press

ENTER.

3. The screen will scroll P2.1, SP2, then the current set point

value.

4. Press

▲ or ▼ keys and adjust second relay set point. You can

adjust it in 0.01 increments from 0.00 to 14.00 pH (-999 to
1000 mV).

5.

Press ENTER to confirm and continue to step three of P2.2.


P2.2: Set relay as high or low set point
Select a low set point to activate controller when your value
undershoots the set point; select a high set point to activate
controller when your value overshoots the set point.
Using both SP1 and SP2, you can select lo/lo, lo/hi, hi/lo or hi/hi set
points.
1.

Key in the password “022” as per procedure in Section 5.

2. Press

▲ key once. The screen will scroll P2.0 and SP2. Press

ENTER twice.

3.

The screen will scroll P2.2, SP2, and LOW (or HIGH).

4. Press

the

▲ or ▼ keys to toggle between LOW and HIGH.

5.

Press ENTER to confirm and continue to step 3 of P2.3.