Thermo Fisher Scientific Alpha pH 190 (In the process of updating the brand guideline & product image) User Manual

Page 14

background image

- 8 -

2. Press

the

▲ or ▼ keys and adjust first relay set point. You can

adjust it in 0.01 increments from 0.00 to 14.00 pH (-999 to
1000 mV).

3. Press ENTER to confirm and return to the measurement

mode.


NOTE: Press ▲ and ▼ keys together (ESCAPE) at anytime, to
return to Measurement mode.


Set Point 2 (SP2) adjustment mode
This lets you adjust the pH (or ORP) value in Set Point 2. If this
value is crossed, the set point relay 2 LED will light.

1. Press the SP2/▲ key. The screen will scroll P2.1, SP2, and

then the current set point value.

2. Press

the

▲ or ▼ keys and adjust second relay set point. You

can adjust it in 0.01 increments from 0.00 to 14.00 pH (-999 to
1000 mV).

3. Press ENTER to confirm and return to the measurement

mode.


NOTE: Press ▲ and ▼ keys together (ESCAPE) at anytime, to
return to Measurement mode.
NOTE: These modes are only for adjusting relay 1 and/or relay 2
set points values. To set the relays as low or high set points or to
set its hysteresis values, you have to make the adjustments from
the Setup mode as in Section 9.3 and Section 9.4.