Appendix 1: Simple Explanation on the Function of Hysteresis

The controller relay activates when the set-
point is reached. In the reverse direction, it
does not de-activate when the value
reaches the set-point. Instead, it continues
to be active till the value reaches the
amount set by the Hysteresis band.

Appendix 2: Factory Defaults

Resetting the controller to factory default settings (See program P7.0) clears
all calibration data and most other setup functions.
The following settings will remain unchanged:
1.

Measurement range

2.

Temperature compensation mode (ATC or MTC)

3.

Temperature offset calibration value, if in ATC mode.

Conductivity Defaults
Conductivity range

Remains unchanged at last selected range

Set point 1

10% full scale / Hyst. 5% full scale / Low

Set point 2

90% full scale / Hyst. 5% full scale / High

Temperature Defaults
Temp. Compensation mode

remains unchanged

MTC mode

reset to 25

o

C if in MTC mode

ATC mode

remains at last calibration if in ATC mode

Temperature Coefficient

2.1 %

Normalization Temperature

25.0

o

C


200

300

1700

1800

RELAY OFF

HYSTERESIS BAND

FORWARD DIRECTION

REVERSE DIRECTION

SP1

SP2

RELAY ON

SP1 Set to LO
SP2 Set to HI

uS