8.2 Setup mode overview
P1.0: Set Point 1

P1.1: select relay 1 set point value

P1.2: select relay 1 as low or high set point

P1.3: set relay 1 hysteresis value


P2.0: Set Point 2

P2.1: select relay 2 set point value

P2.2: select relay 2 as low or high set point

P2.3: set relay 2 hysteresis value


P3.0: Range

P3.1: select measurement range (with the corresponding cell constant value).
Eight measurement ranges are available.


P4.0: Temperature Data

P4.1: select ATC or MTC

P4.2: select temperature coefficient value

P4.3: select normalization temperature


P5.0: Conductivity Calibration points

P5.1: view conductivity value at which calibration was performed


P6.0 Electrode Properties

P6.1: view cell constant value after calibration
P6.2: view temperature offset value after calibration (only if in ATC mode)

P7.0 Reset to Factory Defaults

P7.1: select YES/NO to confirm reset decision