Thermo Fisher Scientific Alpha COND 190 (In the process of updating the brand guideline & product image) User Manual
Page 25

22
8.7 Viewing Conductivity Calibration Data – P5.0 
Program 5 is a “view only” option, which displays the value at which calibration was 
performed. 
 
P5.1: view Calibrated conductivity value (if no calibration performed for this range, 
controller displays ‘- - - - ‘). 
 
These parameters will change each time you recalibrate the controller. 
P5.0
C L
P5.1
CAL
300
Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and 
return to Measurement mode. 
P5.0: Viewing Calibrated Conductivity data 
1. 
Key in the password “022” as per procedure in Section 5.
2.
Press ▲key until screen displays P5.0 and CAL. Press ENTER.
3.
Screen will scroll P5.1, ‘CAL’, then Calibrated value (in uS/cm or mS /cm).
4.
Press ▲and ▼keys together (ESCAPE) twice, to return to measurement 
mode. 
- PCTestr 35 (2 pages)
 - pHScan BNC (3 pages)
 - pHScan 3/3+ (5 pages)
 - pHTestr 1 (3 pages)
 - pHTestr 10/20/30/10 BNC/Spear (2 pages)
 - ORPTestr 10/10 BNC (2 pages)
 - EC/TDS/SaltTestr 11 (4 pages)
 - EC/TDS/SaltTestr (2 pages)
 - ECScan High/Low & TDScan High/Low (9 pages)
 - SaltTestr (2 pages)
 - EcoTestr pH 2 (2 pages)
 - EcoTestr pH 1 (2 pages)
 - EcoTestr EC High (2 pages)
 - EcoTestr EC Low (2 pages)
 - EcoTestr TDS High (2 pages)
 - EcoTestr TDS Low (2 pages)
 - EcoTestr Salt (2 pages)
 - Eutech pH 5/6 Plus & Ion 6 Plus (New version R1.1, SN >797406) (23 pages)
 - Eutech pH 5/6 Plus & Ion 6 Plus (Old version EP6, SN <797406, discontinued) (23 pages)
 - Eutech COND/TDS/Salt 6 Plus (40 pages)
 - Eutech DO 6 Plus (48 pages)
 - EcoScan pH/Ion 5 & 6 (27 pages)
 - EcoScan CON 6 & TDS 6 (56 pages)
 - EcoScan CON 5 & TDS 5 (18 pages)
 - EcoScan Salt 6 (40 pages)
 - EcoScan DO 6 (80 pages)
 - CyberScan pH 10/pH 100 (67 pages)
 - CyberScan pH 11/pH 110 (76 pages)
 - CyberScan CON 10/CON 100/CON 200 (62 pages)
 - CyberScan CON 11/CON 110 (80 pages)
 - CyberScan DO 110 (60 pages)
 - CyberScan PCD 650 (127 pages)
 - CyberScan CON 400/410 (For units manufactured before March 2010, discontinued) (60 pages)
 - CyberScan CON 400 (For units manufactured from March 2010 onwards) (60 pages)
 - CyberScan pH 300/310 (52 pages)
 - CyberScan DO 300 (60 pages)
 - CyberScan PC 300 (72 pages)
 - CyberScan PD 300 (76 pages)
 - CyberScan PC 10 (31 pages)
 - C401 Colorimeter (64 pages)
 - TN100 Turbidimeter (31 pages)
 - RS232C Interface Adapter (9 pages)
 - Thermo Scientific Temp 360 (44 pages)
 - Thermo Scientific Temp 340 (40 pages)
 - Thermo Scientific Temp 300 (32 pages)
 