25

9 RELAYS

The

α

lpha CON 190 features two SPDT non-powered relays; rated for 6A at 110

VAC, 250 VAC maximum. When your process exceeds the set parameters of a
relay set point, the REL 1 or REL 2 indicator will light up.

To set parameters for relay 1 and relay 2, see Setup programs P1.0 and P2.0.

10 TRANSMITTER FUNCTION

If remote data logging is required, a 4-20 mA current loop can be connected. The
current will be proportional to the Conductivity value displayed on the panel and
according to the measurement range selected.

Conductivity Range

Range No., R

Cell Constant

4 mA current

20 mA current

0.00 – 20.00 µS/cm

1

0.1

0.00 µS/cm

20.00 µS/cm

0.0 – 200.0 µS/cm

2

0.1

0.0 µS/cm

200.0 µS/cm

0.0 – 200.0 µS/cm

3

1

0.0 µS/cm

200.0 µS/cm

0 – 2000 µS/cm

4

1

0 µS/cm

2000 µS/cm

0.00 – 10.00 mS/cm

5

1

0.00 mS/cm

10.0 mS/cm

0.00 – 20.00 mS/cm

6

1

0.00 mS/cm

20.0 mS/cm

0.0 – 200.0 mS/cm

7

1

0.0 mS/cm

200.0 mS/cm

0.0 – 200.0 mS/cm

8

10

0.0 mS/cm

200.0 mS/cm



The 4-20 mA current loop can drive a load resistance of no more than 200 Ω.