Wash relay (WASH) sub-function
In this sub-function you set the parameters for the wash relay.
SETUP
SETUP
HOLD
HOLD
4
4
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
4
4
4
SETUP
SETUP
HOLD
HOLD
4
4
ENT
ENT
ENT
ENT
ENT
ENT
ENT
1
2
2
3
3
4
5
1. Select the “WASH” sub-function and then press the ENT key.
2. Enabling/disabling wash function: Press the ▲ or ▼ key to enable (WASH ON) or disable
(WASH OFF) wash function. Press the ENT key to confirm your selection.
− If “WASH OFF” was selected, the transmitter returns “WASH” sub-function.
− If “WASH ON was selected:
3. Selecting the relay status condition: Press the ▲ or ▼ key to choose desired relay
status (de-energised = “DEEN” or energised = “EN”). Press the ENT key to confirm your
selection.
4. Setting the wash interval in hours: Press the ▲ or ▼ key to set the desired wash
interval (setting range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.
5. Setting the wash duration in seconds: Press the ▲ or ▼ key to set the desired wash
duration (setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.
Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the
wash cycle.
Continue with rest of the Setup procedures by pressing ▲ or ▼ key or return to the
Measurement mode by pressing the ▲ and ▼ key (escape) simultaneously.