5calibration mode, 1 entering calibration mode – Thermo Fisher Scientific Alpha COND 2000 (In the process of updating the brand guideline & product image) User Manual
Page 22

Instruction Manual
αlpha CON 2000
16
5
CALIBRATION MODE
You can access the Calibration mode directly from the Measurement mode by pressing the
CAL key and entering the Calibration security code “11”. Calibration mode may also be
accessed via the Setup mode (see section 6.1).
5.1
Entering Calibration mode
MEAS
SETUP
HOLD
µS
4
4
°C
SETUP
HOLD
4
Calibration
See section 5.2
CAL
ENT
ENT
ENT
ENT
ENT
While in Measurement mode press the CAL key (direct access) or ENT key (access via Setup mode).
• If CAL key was pressed: The display prompts you to enter the calibration security
code. Press the ▲ or ▼ key to set the calibration security code to “11”, then press
the ENT key to confirm the calibration security code.
• If ENT key was pressed: The display prompts you to enter the security code. Press
the ▲ or ▼ key to set the security code to “22”. Press the ENT key to confirm the
security code, then press the ▲ or ▼ key to select the calibration sub-function.
The display shows “CAL CON”. To start calibrations refer to section 5.2.
Note: to exit Calibration mode at any time press the ▲ and ▼ keys
simultaneously (escape). The transmitter returns to the Measurement mode and
the old calibration values remain active. The calibration security code automatically
resets from “11” to “000” after the transmitter returns to Measurement mode.