2 temperature compensation (tc) sub-function – Thermo Fisher Scientific Alpha COND 2000 (In the process of updating the brand guideline & product image) User Manual
Page 26

Instruction Manual
αlpha CON 2000
20
6.2
Temperature compensation (TC) sub-function
This sub-function allows you to select the correct temperature compensation for optimum
operations.
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
4
4
4
SETUP
HOLD
4
ATC
°C
SETUP
HOLD
4
ATC
ENT
ENT
ENT
ENT
ENT
1
2
2
3
4
1. Select the “TC” sub-function and then press the ENT key.
2. Selecting type of temperature compensation (TC): Press the ▲ or ▼ key to select the suitable
type of temperature compensation:
− PUR TC = Pure water temperature compensation
Use pure water temperature compensation for applications in the pure water or ultra-
pure water industries.
Note: “PUR TC” should be selected while working with pure water, usually in the 0-
2 µS range.
− LIN TC= Linear temperature compensation
Use linear temperature compensation for all other applications.
Press the ENT key to confirm your selection.
− If linear temperature compensation was selected (LIN TC):
3. Selecting the temperature coefficient value for process liquid (P.TC): Press the ▲
or ▼ key to adjust the temperature coefficient value (setting range: 0 to 10 %, factory
setting: 2.10 %). Press the ENTER key to confirm the temperature coefficient value.
4. Selecting the temperature coefficient value for calibration liquid (C.TC): Press the ▲
or ▼ key to adjust the temperature coefficient value (setting range: 0 to 10 %, factory
setting: 2.10 %). Press the ENT key to confirm the temperature coefficient value.
Continue with additional Setup procedures, or return to the Measurement mode by pressing
the ▲ and ▼ key (escape) simultaneously.