3 view cell constant and calibration factor – Thermo Fisher Scientific Alpha COND 2000 (In the process of updating the brand guideline & product image) User Manual
Page 24

Instruction Manual
αlpha CON 2000
18
Note: If you entered the Calibration mode from the Setup mode, the transmitter will
return to the setup menu.
Note: When calibrating with manual temperature compensation, the transmitter
automatically changes from the preset process temperature to the calibration
temperature. After leaving the Calibration mode, the transmitter switches back to
the process temperature (for setting the calibration temperature and the process
temperature, see section 6.3).
5.3
View cell constant and calibration factor
You can view the calibration data (cell constant & calibration factor) without entering into
calibration mode.
CAL
4
CAL
HOLD
4
CAL
ENTER
ENT
ENTER
ENT
MEAS
µS
4
°C
ENT
1
2
3
4
1. While in Measurement mode press the CAL key.
2. The display prompts you to enter a calibration security code (CCD). Leave the calibration security
code at “000” (view only mode). Press the ENT key.
3. The display shows the cell constant. Press the ENT key.
4. The display shows the calibration factor. An effective electrode should be within the calibration
factor window of 0.60 to 1.40. Press the ENT key to return to the Measurement mode.
Note: If no key is pressed for 30 seconds during the above steps, the transmitter
automatically returns to the Measurement mode.