10 Accessories and Replacement Parts List
Accessory or Replacement Part
Order Code Number
CyberScan TB1000 W Calibration Set with EC-TBDWCALKT
EC-TBDW1000
CyberScan TB1000 IR Calibration Set with EC-TBDIRCALKT
EC-TBDIR1000
Calibration kit set for EC-TBDW1000 (0.02 NTU, 10.0 NTU, 1000 NTU)
EC-TBDWCALKT
Calibration kit set for EC-TBDIR1000 (0.02 NTU, 10.0 NTU, 1000 NTU)
EC-TBDIRCALKT
Lamp Module – Tungsten Filament
EC-TNGSTNLMP
Pour Through Assembly
EC-PORTHASSY
Sample Cuvettes – 3 pack
EC-TBDCUV03KT
Cuvette Maintenance Kit
EC-CUVMNTCKT
Internal Batteries (set of 2)
EC-BATT3032
RS232 Cable for Serial Printer
EC-MCR0100CBL
120 VAC/12 VDC Power Adapter (50/60 Hz)
EC-120ADA800
220 VAC/12 VDC Power Adapter (50/60 Hz)
EC-220ADA800
10.1 Glossary
Formazin Turbidity Units (FTU): see Nephelometric Turbidity Units
Indexing a Cuvette: The United States Environmental Protection Agency (US EPA) recommends that cuvettes
used for turbidimeter calibration or sample measurement be indexed. To index a cuvette with a sample in it,
slowly rotate the cuvette throughout one complete revolution (360°). While rotating the sample cuvette, observe
the display and locate the position that the cuvette is in which provides the lowest turbidity reading. This position
is the indexed position of the cuvette.
Nephelometric Turbidity Units (NTU): Unit of measure used when comparing the light scattered by a liquid media
to the light scattered by a known concentration of Formazin Polymer. This unit of measure is recognized as a
measure of the optical clarity of an aqueous sample. NTU is the accepted unit of measurement for turbidity.
Turbidity: 1) A measure of the attenuation of a radiant flux as it passes through a liquid media. 2) Optical clarity of
a liquid, 3) a phenomena caused by the presence of undissolved matter in liquid media.