Thermo Fisher Scientific CyberScan PD 300 User Manual
Page 61
![background image](/manuals/370621/61/background.png)
Instruction Manual
PD 300
Close
Open
Installation tool
Insert installation
tool into slots on
membrane lock.
Then unscrew
membrane lock
from membrane
cap
Figure 44: Use tool to take out (or
put in) membrane
10. Using the installation tool, screw the membrane
lock back into the cap. Tighten the lock firmly over
the membrane and O-ring, but do not over tighten.
11. Inspect the membrane for wrinkles. If wrinkles
exist, remove the membrane and repeat steps 8 –
11.
12. Fill the membrane cap with water and inspect the
bottom for leaks. If water drops are leaking from
the membrane, re-seal the membrane on the O-
ring (repeat steps 8 – 11, for membrane
replacement only).
13. If the assembly is leak-free, empty the water and
fill the membrane cap with electrolyte to the brim.
14. Tap the side of the housing gently to remove any
air bubble that may be sticking to the membrane.
15. Screw the cap onto the probe. Excess electrolyte
will drain out.
16. Replace probe guard.
17. Calibrate the probe (see section 4) after the %
saturation readings have stabilised.
NOTE
Membranes can only be used once. When a membrane
cap is screwed onto the probe, the membrane is
stretched by the cathode. If the same O-ring and the
membrane is used a second time it will not fit
perfectly onto the cathode. This will result is erratic
readings.
Figure 45 : Membrane cap housing
with O-ring & membrane
56