H 10 – Thermo Fisher Scientific CyberScan PD 300 User Manual

Page 54

background image

Instruction Manual

PD 300

8.12

P7.0 [6.0] rSt: Resetting to Factory Default Setting (DO)

This program lets you reset all DO (% saturation and mg/L) parameters to factory default
settings. This clears all calibration data and any other DO setup functions you might have
changed.

The following settings will remain as you have set them:

Temperature unit of measure (°C or °F)

The temperature offset calibration value

All DO calibration data and parameters

There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.

From measurement mode

Figure 41: Reset to factory

defaults

1.

Press SETUP key to enter SETUP mode.

2. Press

▲ or ▼ keys to scroll through subgroups

until you view parameter “rSt” in the upper display.

3.

Press the ENTER key.

4. Press

▲ or ▼ keys to toggle between NO and

YES.

NO retains current memory.

YES clears all memory

5. Press the ENTER key to confirm selection and to

return to measurement mode.

49