8

I

Mains Switch On

O

Mains Switch Off

Replacement cable
Should the mains lead need replacement, a cable of 1mm

2

of harmonized code H05VV-F connected

to an IEC320 plug should be used. IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN.

ENVIRONMENTAL CONDITIONS

The Hybrigene is designed operate under the following conditions:

Indoor use

Ambient temperature range +5ºC to +40ºC

Altitude to 2000m

Relative humidity up to 80%

Mains supply fluctuations not exceeding 10%

Over voltage category II IEC 60364-4-443

Pollution degree 2

Use with a minimum distance all-around of 200mm from walls or other items

Note: The control specifications are quoted at an ambient temperature of 20ºC. The specification
may deteriorate outside an ambient temperature of between 18ºC and 30ºC.

The instrument has been tested for radio frequency interference and is certified under EN61326.

GUARANTEE

The instrument is guaranteed against any defect in material or workmanship for the period specified

on the enclosed guarantee card. This period is effective from the date of purchase; within this period

all defective parts will be replaced free of charge provided that the defect is not the result of an

accident, misuse or negligence. Servicing under this guarantee should be obtained from the supplier

of the instrument.

This manual has been prepared for the convenience of Techne’s customers and nothing in this

manual shall be taken as a warranty, condition or representation concerning the description,

merchantability, fitness for purpose or otherwise of the unit or components.

Notwithstanding the description and specification(s) of the instruments contained in the operator’s

manual, Techne reserves the right to make such changes as it sees fit to the instruments or to any of

the components.