8 command #138 read analog output function – KROHNE UFC 500 HART User Manual
Page 22
HARTô Smart Communications Protocol
UFC500 Command Specification Rev: 2.0 Release Date: 9 Jun 2001
© 2001 KROHNE Mefltechnik D-47058 Duisburg
Page
22
8.8
Command #138
Read Analog Output Function
Reads the function of current/pulse output.
REQUEST DATA BYTES
#0
ANALOG
OUTPUT
NUMBER
CODE
RESPONSE DATA BYTES
#0
#1
ANALOG
ANALOG
OUTPUT
OUTPUT
NUMBER
FUNCTION
CODE
CODE
Data Byte #0
Analog Output Number Code, equals м1о for current output and м2о - for pulse output
Data Byte #1
Analog Output Function Code, depends on the output type:
For current output:
0 - Off.
1 - Direction indication.
2 - 1 Dir, output traces flowrate measurements only in one (positive) direction.
3 - Forward/reverse flow, the complete flowrate dynamic range (PV Upper Range Value +
Reverse Range) is mapped onto the range of the output current. Refer also to the instrument
Operating Manual and to Commands #140, Read Current Output Parameters, #158, Read
Transmitter Variable Range Values.
4 - 2 Dir, output traces flowrate measurements in both directions, for every direction the
corresponding flowrate dynamic range (PV Upper Range Value for positive direction and
Reverse Range - for negative) is mapped onto the range of the output current. Refer also to
the instrument Operating Manual and to Commands #140, Read Current Output Parameters,
#158, Read Transmitter Variable Range Values.
5 - Transit time.
For frequency/pulse output:
0 - Off.
1 - Error indication.
2 - 1 Dir, output traces flowrate measurements only in one (positive) direction.
3 - 2 Dir, output traces flowrate measurements in both directions.
4 - Transit time.
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1
Undefined
2
Invalid Selection (of Analog Output)
3-4
Undefined
5
Too Few Data Bytes Received
6
Local Device User
7-127
Undefined