8.6

Command #135

Write Analog Output Low Flow Cutoff Control And Values

Writes low flow cutoff parameters for any of the two outputs.

REQUEST DATA BYTES

#0

#1

#2

#3

ANALOG CUTOFF

CUTOFF

CUTOFF

OUTPUT

CONTROL

ON

OFF

NUMBER

VALUE

VALUE

CODE

RESPONSE DATA BYTES

#0

#1

#2

#3

ANALOG CUTOFF

CUTOFF

CUTOFF

OUTPUT

CONTROL

ON

OFF

NUMBER

VALUE

VALUE

CODE

Data Byte #0

Analog Output Number Code, equals м1о for current output and м2о - for pulse output.

Data Byte #1

Cutoff control variable:
0 - enables cutoff option,
1 - disables cutoff option.
Even when the cutoff option is switched from On to Off, the rest data in request packet is
processed and the On/Off Values will be altered in transmitter EEPROM.

Data Byte #2

Cutoff On Value, in % of the Primary Variable Upper Range Value. This byte is meaningful
even if Data Byte #0 = 0. Value must belong to interval 1..19%.

Data Byte #3

Cutoff Off Value, in % of the Primary Variable Upper Range Value. This byte is meaningful
even if Data Byte #0 = 0. Value must belong to interval 2..20%.

Refer also to comments for the previous command.

NOTE: Low flow cutoff settings can be altered only when the function of corresponding output is related to flowrate
measurements. Refer to Command #138, Read Analog Output Function.

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1 Undefined
2

Invalid Selection (of Analog Output)

3-4

Undefined

5

Too Few Data Bytes Received

6

Local Device User

7

In Write Protect Mode

8

Undefined

9

Off Value too High

10

Off Value too Low

11

On Value too High

12

On Value too Low

13

On Value not Lower than Off Value

14

Undefined

15

Invalid Cutoff Control or Output Function

16-127

Undefined