6.6 Guidelines for maximum operating pressure

Notes:

• Ensure that the meter is used within its operating limits
• All hygienic process connections have a maximum operating rating of 10 barg

at 130°C / 145 psig at 266°F

Linear de-rating of PED certified secondary containment

Pressure / temperature de-rating, all meter sizes in metric (flanged connections as

Pressure / temperature de-rating, all meter sizes in metric (flanged connections as

Pressure / temperature de-rating, all meter sizes in metric (flanged connections as

Pressure / temperature de-rating, all meter sizes in metric (flanged connections as
per EN 1092-1:2007)

per EN 1092-1:2007)

per EN 1092-1:2007)

per EN 1092-1:2007)

X temperature [°C]
Y pressure [barg]

1 Measuring tube (UNS S32760) PED certification
2 Measuring tube (UNS S31803) PED certification
3 Measuring tube (UNS S31803 / S32760) FM certification
4 Measuring tube (UNS S31803) CRN certification

Outer cylinder material

-45°C

20°C

130°C

304 / L or 316 / L

40 barg

40 barg

32 barg

UNS S31803

150 barg

150 barg

100 barg

MA_OPTIMASS_2400_en_140515_4003533101_R01.book Page 38 Thursday, July 31, 2014 5:08 PM