KROHNE ECOFLUX IFS 1000 EN User Manual
Page 8

8
Nominal size
measuring
tube and
pipe flanges
Pressure
rating or
flange class
of pipe flanges
Max.
allowable
operating
pressure
bar
psig
1)
For meter sizes DN 10 and
3
/
8
”, use pipe flanges DN 15 or
1
/
2
”.
6 Torques
Max.
torques
stud bolts
Nm
ft x lbf
. . . DIN 2501 (BS 4504)
DN 10-25 1)
PN 16/PN 40
≤
16
≤
230
0
16
11.6
DN 40
PN 16/PN 40
≤
16
≤
230
0
25
18.1
DN 50
PN 16/PN 40
≤
16
≤
230
0
45
32.5
DN 80
PN 16/PN 40
≤
16
≤
230
0
25
18.1
DN 100
PN 16/PN 40
≤
16
≤
230
0
33
23.9
DN 150
PN 16/PN 40
≤
16
≤
230
0
82
59.3
. . . ANSI B 16.5
3
/
8
”-
1
/
2
” 1)
150/300 lb
≤
16
≤
230
0
16
11.6
1”
150/300 lb
≤
16
≤
230
0
15
10.8
1
1
/
2
”
150/300 lb
≤
16
≤
230
0
25
18.1
2”
150/300 lb
≤
16
≤
230
0
45
32.5
3”
150 lb
≤
16
≤
230
0
56
40.5
300 lb
0
28
20.5
4”
150/300 lb
≤
16
≤
230
0
36
26.0
6”
150 lb
≤
16
≤
230
100
72.3
300 lb
0
66
47.7
. . . JIS
DN 10-15 1)
10 K
≤
0
7
≤
100
0
16
11.6
20 K
≤
14
≤
200
0
16
11.6
DN 25
10 K
≤
0
7
≤
100
0
16
11.6
20 K
≤
14
≤
200
0
19
13.7
DN 40
10 K
≤
0
7
≤
100
0
20
14.5
20 K
≤
14
≤
200
0
17
12.3
DN 50
10 K
≤
0
7
≤
100
0
35
25.3
20 K
≤
14
≤
200
0
24
17.4
DN 80
10 K
≤
0
7
≤
100
0
20
14.5
20 K
≤
14
≤
200
0
30
21.7
DN 100
10 K
≤
0
7
≤
100
0
26
18.8
20 K
≤
14
≤
200
0
89
28.9
DN 150
10 K
≤
0
7
≤
100
0
65
47.0
20 K
≤
14
≤
200
0
89
64.3
7. Grounding
•
Warning: All flowmeters must be properly grounded to avoid personnel shock hazard.
•
The ground conductor should not transmit any interference voltages, therefore do not ground
any other electrical devices together with this conductor.
Separate primary head IFS 1000 F with terminal box
•
A functional ground FE must always be connected.
•
Signal converters with a field power supply of > 125 mA / > 60 V for the primary heads
should not be operated with the IFS 1000 F.