KROHNE ECOFLUX IFS 1000 EN User Manual

Page 8

background image

8

Nominal size
measuring
tube and
pipe flanges

Pressure
rating or
flange class
of pipe flanges

Max.
allowable
operating
pressure

bar

psig

1)

For meter sizes DN 10 and

3

/

8

”, use pipe flanges DN 15 or

1

/

2

”.

6 Torques

Max.
torques
stud bolts

Nm

ft x lbf

. . . DIN 2501 (BS 4504)
DN 10-25 1)

PN 16/PN 40

16

230

0

16

11.6

DN 40

PN 16/PN 40

16

230

0

25

18.1

DN 50

PN 16/PN 40

16

230

0

45

32.5

DN 80

PN 16/PN 40

16

230

0

25

18.1

DN 100

PN 16/PN 40

16

230

0

33

23.9

DN 150

PN 16/PN 40

16

230

0

82

59.3

. . . ANSI B 16.5

3

/

8

”-

1

/

2

1)

150/300 lb

16

230

0

16

11.6

1”

150/300 lb

16

230

0

15

10.8

1

1

/

2

150/300 lb

16

230

0

25

18.1

2”

150/300 lb

16

230

0

45

32.5

3”

150 lb

16

230

0

56

40.5

300 lb

0

28

20.5

4”

150/300 lb

16

230

0

36

26.0

6”

150 lb

16

230

100

72.3

300 lb

0

66

47.7

. . . JIS
DN 10-15 1)

10 K

0

7

100

0

16

11.6

20 K

14

200

0

16

11.6

DN 25

10 K

0

7

100

0

16

11.6

20 K

14

200

0

19

13.7

DN 40

10 K

0

7

100

0

20

14.5

20 K

14

200

0

17

12.3

DN 50

10 K

0

7

100

0

35

25.3

20 K

14

200

0

24

17.4

DN 80

10 K

0

7

100

0

20

14.5

20 K

14

200

0

30

21.7

DN 100

10 K

0

7

100

0

26

18.8

20 K

14

200

0

89

28.9

DN 150

10 K

0

7

100

0

65

47.0

20 K

14

200

0

89

64.3

7. Grounding

Warning: All flowmeters must be properly grounded to avoid personnel shock hazard.

The ground conductor should not transmit any interference voltages, therefore do not ground
any other electrical devices together with this conductor.

Separate primary head IFS 1000 F with terminal box

A functional ground FE must always be connected.

Signal converters with a field power supply of > 125 mA / > 60 V for the primary heads
should not be operated with the IFS 1000 F.