8.3 Return / maintenance form

Customer details

Device details

If yes please fill out the table with following check points.

Before you begin, configure the TT 51 for RTD measurement 3-wire connection on both
channels. Select measuring range 0...+100°C, output – dedicated dynamic variable Ch1 select
sensor break and sensor short circuit to downscale, therefore to upscale value.

Select even sensor backup.

Proof test check points

Company:

Address:

Contact person:

Telephone:

Fax:

Email:

Product ID:

Serial no.:

Reason for the return / maintenance:

Have you performed the proof test on the product?

Yes

No

Step

Description

Yes

No

Comments

1

Connect the selected sensors on Ch 1 and Ch2 and

check for the output range value is within measuring

range.

2

Simulate sensor break for each single wire (on

terminals 1...5) and check the output value (≥21 mA) /

(≤3.6 mA).

3

Simulate sensor short between 1...5 terminals and

check the output value (≥21 mA) / (≤3.6 mA).

4

Simulate sensor break or sensor short (one error at a

time) for sensor connected on Ch 1. Check if the

transmitter will switch automatically over to

measuring on Ch 2.