beautypg.com

Periodic checks / proof tests – KROHNE TT 51 SERIES EN User Manual

Page 12

background image

6

PERIODIC CHECKS / PROOF TESTS

12

TT 51 SERIES

www.krohne.com

09/2010 - 4000869801 - AD TT 51 SIL R01 en

Proof test check points

• Repeat configurations points 7...8.2 of the proof test configuration and change to down scale

error value (≤3.6 mA).

• Repeat all check points (to be sure the transmitter is not stuck in some of conditions).

8

Select device information tab. Specify a mounting date in tag field.

8.1

Describe the proof test in the description field and date of the test.

8.2

Specify any other information in the message field.

Step

Description

Step

Description

Yes

No

Comments

1

Connect the selected sensors on Ch 1 and Ch 2 and

check for the output range values.

2

Simulate sensor break for each single wire and check

the output value (≥21 mA).

3

Simulate sensor short between 1...5 terminals and

check the output value (≥21 mA).

4

Simulate sensor break or sensor short (one error at a

time) for sensor connected on Ch 1. Check if the

transmitter will switch automatically over to

measuring on Ch 2.

.book Page 12 Thursday, September 9, 2010 4:19 PM