ELECTRICAL CONNECTIONS

4

15

DK32

www.krohne.com

04/2011 - 4000898801 MA DK32 Air purge set R02 en

4.3 Setting the limit switch

Set the pointers to the desired limit values as a min. contact 1 and max. contact 2 using a slip
coupling along the scale.

Adjusting the limit switch with reed contact:

• Slacken nut 3
• Set reed cartridge to the required value
• Secure with nut 3

Comment

The reed contact is actuated directly by the float magnet. The desired operating point can only be
determined in measuring mode. A reference to the scale and/or pointer cannot be established.

Connection three-wire reed contact

Figure 4-2: Electrical connection of reed contact limit switch

Strand colours for flowmeters with preassembled cables:

1 Silicone-insulated wire - yellow/green / FEP-insulated wire - red
2 Silicone-insulated wire - brown / FEP-insulated wire - brown
3 Silicone-insulated wire - blue / FEP-insulated wire - blue

Figure 4-3: Limit switch settings