beautypg.com

Technical data – KROHNE OPTIFLUX 2000 EN User Manual

Page 38

background image

6

TECHNICAL DATA

38

OPTIFLUX 2000

www.krohne.com

4/2014 - 4000839504 - HB OPTIFLUX 2000 R04 en

Verification to MI-001

INFORMATION!
Verification to MI-001 and to OIML R49 class 2 is carried out at the following values for R, Q1, Q2
and Q3. Verification to OIML R49 class 1 and at other values for R and Q3 available on request.

DN

Span (R)

Q3 / Q1

Flow rate [m

3

/h]

Q1

Q2

Q3

25

80

0.05

0.08

4

32

80

0.125

0.20

10

40

80

0.125

0.20

10

50

80

0.2

0.32

16

65

80

0.3125

0.50

25

80

80

0.5

0.80

40

100

80

0.7875

1.26

63

125

80

1.250

2.00

100

150

80

2.0

3.2

160

200

80

3.125

5.0

250

250

80

5.0

8.0

400

350

80

5.0

8.0

400

300

80

7.875

12.6

630

350

80

12.5

20.0

1000

400

80

12.5

20.0

1000

450

80

12.5

20

1000

500

80

20.0

32

1600

600

80

50.0

80

4000

700

80

125

200

10000

800

80

125

200

10000

900

80

200

320

16000

1000

80

200

320

16000

1200

80

200

320

16000

1600

80

312.5

500

25000