Technical data – KROHNE OPTIFLUX 2000 EN User Manual

Page 29

background image

TECHNICAL DATA

6

29

OPTIFLUX 2000

www.krohne.com

4/2014 - 4000839504 - HB OPTIFLUX 2000 R04 en

Measuring accuracy

Reference conditions

Medium: water

Temperature: +10...+30°C / +50...+86°F

Operating pressure: 1 bar / 14.5 psi

Inlet section ≥ 5 DN

Electrical conductivity: ≥ 300 μS/cm

Maximum measuring

error

IFC 050: 0.5% of the measured value above 0.5 m/s; below 0.5 m/s, deviation
±

2.5 mm/s

IFC 100: down to 0.3% of the measured value ±1 mm/s

IFC 300: down to 0.2% of the measured value ±1 mm/s

The maximum measuring error depends on the installation conditions.

For detailed information refer to

Measuring accuracy

on page 39.

Repeatability

±

0.1% of the measured value, minimum 1 mm/s

Calibration / Verification

Standard:

Standard:

Standard:

Standard:

2 point calibration by a direct volume comparison.

Optional:

Optional:

Optional:

Optional:

Verification to Measurement Instrument Directive (MID), Annex MI-001.

Standard: Verification at Ratio (Q3/Q1) = 80, Q3 ≥ 2 m/s

Optional: Verification at Ratio (Q3/Q1) > 80 on request

Only in combination with the IFC 300 signal converter.

MID Annex MI-001

(Directive 2004/22/EC)

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

Only in combination with the IFC 300 signal converter.

Diameter range: DN25...1600

Forward and reverse (bi-directional) flow

Liquid temperature range: +0.1°C / 50°C

For detailed information refer to

Legal metrology

on page 34.

OIML R49

Certificate of conformity to OIML R49

Certificate of conformity to OIML R49

Certificate of conformity to OIML R49

Certificate of conformity to OIML R49

Only in combination with the IFC 300 signal converter.

Diameter range

Class 1:DN65...1600

Class 2: DN25...50

Forward and reverse (bi-directional) flow

Liquid temperature range: +0.1°C / 50°C

For detailed information refer to

Legal metrology

on page 34.