background image

4

ELECTRICAL CONNECTIONS

24

OPTIFLUX 2000

www.krohne.com

4/2014 - 4000839504 - HB OPTIFLUX 2000 R04 en

4.3 Virtual reference for IFC 300 (C, W and F version)

The virtual reference option on the IFC 300 flow converter provides complete isolation of the
measurement circuit.

Benefits of virtual reference:

• Grounding rings or grounding electrodes can be omitted.
• Safety increases by reducing the number of potential leakage points.
• The installation of the flowmeters is much easier.

Minimum requirements:

• Size: ≥ DN10
• Electrical conductivity: ≥ 200 µS/cm
• Electrode cable: max. 50 m / 164 ft, type DS

4.4 Connection diagrams

Figure 4-3: Virtual reference

INFORMATION!
For the connection diagrams please refer to the documentation of the applicable signal
converter.