Technical data – KROHNE OPTIFLUX 2000 EN User Manual
Page 32
6
TECHNICAL DATA
32
OPTIFLUX 2000
www.krohne.com
4/2014 - 4000839504 - HB OPTIFLUX 2000 R04 en
Grounding rings
Standard: stainless steel
Option: Hastelloy
®
C, titanium, tantalum
Grounding rings can be omitted with virtual reference option for the
IFC 300 signal converter.
Reference electrode
(optional)
Standard: Hastelloy
®
C
Option: stainless steel, titanium
Other materials on request
Process connections
Flange
Flange
Flange
Flange
EN 1092-1
DN25...3000 in PN 2.5...40
ASME
1…24" in 150 & 300 lb RF
JIS
DN25…1000 in 10...20 K
AWWA
DN700...2000 in 6...10 bar
Design of gasket surface
RF
Other sizes or pressure ratings on request.
Electrical connections
For full detail; see the relevant documentation of the signal converter
Signal cable
Signal cable
Signal cable
Signal cable (remote versions only)
Type A (DS)
In combination with the IFC 050, IFC 100 and IFC 300 signal converter
In combination with the IFC 050, IFC 100 and IFC 300 signal converter
In combination with the IFC 050, IFC 100 and IFC 300 signal converter
In combination with the IFC 050, IFC 100 and IFC 300 signal converter
Standard cable, double shielded.
Max. length: 600 m / 1950 ft
(depends on electrical conductivity and measuring sensor)
Type B (BTS)
Only in combination with the IFC 300 signal converter
Only in combination with the IFC 300 signal converter
Only in combination with the IFC 300 signal converter
Only in combination with the IFC 300 signal converter
Optional cable, triple shielded.
Max. length: 600 m / 1950 ft
(depends on electrical conductivity and measuring sensor)
I/O
For full details of I/O options, including data streams and protocols, see
technical datasheet of the relevant signal converter.
Approvals and certificates
CE
CE
CE
CE
This device fulfils the statutory requirements of the EC directives. The
manufacturer certifies successful testing of the product by applying the CE
mark.
Electromagnetic
compatibility
Directive: 2004/108/EC, NAMUR NE21/04
Harmonized standard: EN 61326-1 : 2006
Low voltage directive
Directive: 2006/95/EC
Harmonized standard: EN 61010 : 2010
Pressure equipment
directive
Directive: 97/23/EC
Category I, II, III or SEP
Fluid group 1
Production module H