beautypg.com

Technical data, 2 technical data – KROHNE OPTISENS PH 8590 User Manual

Page 41

background image

TECHNICAL DATA

6

41

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

6.2 Technical data

pH electrodes

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

OPTISENS

PH 8590

Measuring system

Measuring principle

Potentiometric

Measuring range

0…14 pH

Design

Construction

CPVC Body with Tuff Tip®

Shaft diameter

20 mm / 0.79"

insertion length

23 mm / 0.91" or 44 mm / 1.74"

Process Connection

3/4" MNPT

Temperature sensor

Pt 100

Sensor Cap

attached cable or 4-pin connector

Type of diaphragm

Ceramic

Measuring accuracy

Reference conditions

Medium: water

Temperature: 20°C / 68°F

Pressure: max. 1 bar / 14.5 psi (absolute)

Maximum measuring

error

pH: 0.2% full scale

Temperature: 1.0% full scale

Repeatability

0.2% full scale

Resolution

0.1 (or 0.01 in extended mode)

Long-term stability

24 hours: tested within accuracy definition

Temperature drift

Tested within accuracy definition

Cable length variation

Tested within accuracy definition

.book Page 41 Monday, March 19, 2012 2:01 PM