3

INSTALLATION

18

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

3.8 Installing an external temperature sensor

Since the slope of the pH electrode is temperature-dependent, a temperature compensation
makes sense if the temperature of your measuring medium varies.

If the pH electrode does not have an internal Pt100/1000 temperature sensor, you should use an
external temperature sensor. There are various different possibilities of adding an external
temperature sensor to your pH measurement installation. They can not be explain in detail in
this document. Nevertheless, you should follow the general guidelines.

General guidelines for external temperature sensor installation

• The temperature sensor should be calibrated using a reference thermometer. For further

information refer to

Temperature compensation on page 25.

• The temperature should be measured close to the pH electrode in the same medium to avoid

temperature differences. For example, surface water may have a different temperature than
deeper water.

Suitable external temperature sensors are

• Pt100 in 2-wire or 3-wire connection
• Pt1000 in 2-wire or 3-wire connection