Operation, 4 troubleshooting – KROHNE OPTISENS PH 8590 User Manual
Page 34
4
OPERATION
34
OPTISENS PH 8590
www.krohne.com
03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en
4.3.3 Calibration log
Accessing the calibtration log
• You are on the data level and you see the calibration history. With the help of
or
you can
scroll through the different entries.
• If you want to return to the measuring mode press ^
^
^
^ several times until you reach this mode.
4.4 Troubleshooting
INFORMATION!
In order to show the history of the calibrations, the converter has a calibration logbook function.
Up to 64 entries of the calibration history are stored including date and time.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu test
test
test
test is highlighted.
MAIN MENU
A quick setup
> B test
> B test
> B test
> B test
C setup
D service
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "test" and "B1" appears,
beneath the submenu sim.process input A
sim.process input A
sim.process input A
sim.process input A is highlighted.
Press or until the submenu logbooks
logbooks
logbooks
logbooks is highlighted.
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display "logbooks" and
"B1" appears, beneath the submenu status log
status log
status log
status log is highlighted.
Press
or
until the submenu calibration log
calibration log
calibration log
calibration log is highlighted.
Press >
>>> to enter the chosen menu.
Problem
Possible cause
Remedy
The pH electrode
does not deliver a
signal.
Mechanical damage of the
glass bulb, e.g. small
cracks. This will shortcut
the ion exchange and
deliver a pH7 reading.
Exchange electrode.
Moisture inside the
electrode connector. This
will shortcut the signal
between pH glass half cell
and reference half cell
and deliver a reading of
pH7.
Clean the connector (electrode/cable) with pure water
and dry with air (e.g. hair blower).
The pH electrode
delivers an
unstable signal.
The diaphragm in the
reference half cell does
not provide good contact
to the process medium
due to drying up or
coatings.
•
Clean the diaphragm with hot soap or acid using a
soft tissue (details on page 35).
•
Submerge electrode in water and increase the
temperature to 50...60°C / 122...140°F.
•
Submerge electrode in 3 molar KCl solution at
ambient temperature. The decrease in temperature
will cause the reference half cell to suck in KCl
solution through the diaphragm and regenerate the
diaphragms functionality.
.book Page 34 Monday, March 19, 2012 2:01 PM