beautypg.com

Installation, 3 installation procedure – KROHNE OPTISENS PH 8590 User Manual

Page 12

background image

3

INSTALLATION

12

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

Storing the electrode in the provided plastic tube

• Take the protection cap from the packaging.
• Fill the protection cap with 3 molar KCl solution.
• Insert the electrode tip into the protection cap until it is fully covered with KCI solution.
• Store the electrode in its original packaging.

3.3 Installation procedure

Because a new pH electrode needs to be calibrated before it is installed into its final measuring
location, it is important to follow the installation order:

1. Unpack the electrode.
2. Connect the electrode to the signal converter.
3. Calibrate the electrode.
4. Install the electrode into its final measuring location.

The required steps are explained in the following sections.

Figure 3-1: pH electrode in its original packaging

1 Protection cap filled with 3 molar KCl Solution
2 pH electrode
3 attached cable or 4-pin connector

.book Page 12 Monday, March 19, 2012 2:01 PM