4

OPERATION

26

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

After starting-up the signal converter, the measuring screen appears. This is the standard
screen which is displayed automatically in the normal operating mode. If you are in this mode
and you want to adjust the temperature compensation, you have to perform the following steps:

Step 1: activating the temperature compensation for measurement

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu setup

setup

setup

setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

> C setup

> C setup

> C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,

beneath the submenu process input A

process input A

process input A

process input A is highlighted.

Press

or

to select process input A

process input A

process input A

process input A or process input B

process input B

process input B

process input B. Choose process input A or B

where ever pH is configurated.

Press >>>> to enter the chosen menu.

You are on the second submenu level. The submenu parameter

parameter

parameter

parameter is highlighted.

Press >>>> to enter the chosen menu.

You are on the third submenu level. In the upper line of the display "parameters"

appears

Press

or

to select pH

pH

pH

pH

Press ^

^

^

^ to confirm the entered value

You are on the second submenu level. The submenu parameter

parameter

parameter

parameter is highlighted.

Press

or

to select temperature

temperature

temperature

temperature

Press >>>> to enter the chosen menu.

You are on the third submenu level. In the upper line of the display "temperature"

and "C1.15.1" or "C2.15.1" appears, beneath the option probe

probe

probe

probe is highlighted

Press

or

to select temp.comp

temp.comp

temp.comp

temp.comp

Press >>>> to enter the chosen menu.

You are on the fourth submenu level. In the upper line of the display

"temp.comp" appears

Now you can setup the temperature compensation

Press

or

to choose between automatic or off

Press ^

^

^

^ to confirm the entered value

You are on the third submenu level. If you have chosen the option automatic, you

can select the type of temperature compensation now.

Press

or

to choose probe

Press >>>> to enter the chosen menu.

You are on the fourth submenu level. Press

or

to choose between

Pt100, Pt1000, manual or process input A.

Press ^

^

^

^ to confirm the entered value

Press ^

^

^

^ several time until you reach the measuring mode again. Choose

yes

yes

yes

yes to safe and confirm your selection

INFORMATION!
When ordering the dual channel version, the interface "Pos A" and "Pos B" are populated. A
separate temperature sensor has to be connected to "Pos A" and configurated on process input

process input

process input

process input

A

A

A

A. To connect a electrode with integrated Pt100/Pt1000 the temperature sensor has to be
connected on the same terminal block as the electrode.