beautypg.com

Installation, 7 installing the sensor – KROHNE OPTISENS PH 8590 User Manual

Page 17

background image

INSTALLATION

3

17

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

3.7 Installing the sensor

3.7.1 General installation instructions

The electrode tip must always have full contact with the measuring medium.

The mounting position of the electrode should not deviate more than 75° from vertical position
(electrode tip pointing downwards). Doing otherwise might cause internal air bubbles to float
into the electrode's glass tip. This would interrupt the electrical contact between the inner buffer
solution and the glass surface.

Figure 3-7: Installation requirements

1 Measuring medium
2 Maximum deviation of 75° from vertical position

.book Page 17 Monday, March 19, 2012 2:01 PM