C1.15.5

C2.15.5

temp. comp.

Menu for activating the temperature compensation parameters for the

measurement.

Options:

Linear: linear temperature compensation.

Off: temperature compensation is disabled.

C1.16

C1.16

C1.16

C1.16

C2.16

C2.16

C2.16

C2.16

REDOX cal.

menu item for calibrating the ORP sensor

C1.16.1

C2.16.1

prepare calibration

view actual redox value

C1.16.2

C2.16.2

ref. value Redox

Enter the reference value of the probe in mV

C1.16.3

C2.16.3

start calibration

Start calibration procedure.

C1.16.4

C2.16.4

stored value

View stored value of calibration.

C1.17

C1.17

C1.17

C1.17

C2.17

C2.17

C2.17

C2.17

pH cal.

menu item for calibration the pH sensor

C1.17.1

C2.17.1

temp.comp

Menu for activating the temperature compensation parameters for the

calibration

Options:

off: temperature measurement is disabled

manual: temperature value has to be entered manually

automatic: temperature measurement is performed as configured

C1.17.2

C2.17.2

start calibration in buffer

1

Start calibration procedure.

C1.17.3

C2.17.3

stored value

View stored value of calibration.

C1.17.4

C2.17.4

reference

Enter the reference value of the probe in pH

C1.17.6

C2.17.6

start calibration in buffer

2

Start calibration procedure.

C1.17.7

C2.17.7

stored value

View stored value of calibration

C1.17.8

C2.17.8

reference

Enter the reference value of the probe in pH

CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.