KROHNE CORIMASS MFC 85 Interface EN User Manual
Page 19

19
8.
Format of the Data Blocks
8.1
Measurement Block, Function Code (FKT) = OO Hex
Offset
Variable/Format
Length
Meaning
0
drive_level
(int, LSB first)
2
Installation factor
2
mass_flow_rate_LPF
(float, inversed IEEE 754)
4
Mass flow in g/sec
6
master_total
(double, LSB first)
8
Mass totalizer in g
14/0E hex
volume_total
(float, inversed IEEE 754)
4
Volume totalizer in cm
3
18/12 hex
tube_temperature
(int, LSB first)
2
Temperature in
°
C
×
10
20/14 hex
strain
(int, LSB first)
2
Strain gauge in
Ω
×
20
22/16 hex
frequency
(float, inversed IEEE 754)
4
Resonant frequency in Hz
26/1A
density_LPF
(float, inversed IEEE 754)
4
Fluid density in g/cm
3
30/1E hex
zeroadj_flow_LPF
(float, inversed IEEE 754)
4
Mass flow in g/sec. Valid during zero
calibrations
34/22 hex
phase
(float, inversed IEEE 754)
4
Raw sensor phase in radians. Used for
debug purposes.
38/26 hex
percentage_by_vol
float, inversed IEEE 754)
4
Valid only for General Concentration
option.
Units: 0.01% (thus a value of 1.0
corresponds to 100%)
42/2A hex
percentage_by_mass
(float, inversed IEEE 754)
4
Valid if any concentration option is
active.
Units: 0.01 of concentration unit related
to option.
i.e. a value of 1.0 corresponds to:
100% for General Concentration
100
°
Brix for Brix Concentration
100
°
Baumé for Baumé Concentration
46/2E hex
solid_flow_rate
(float, inversed IEEE 754)
4
Valid if any Concentration option is
active in g/sec.
50/32 hex
sum_angle
(float, inversed IEEE 754)
4
Used for debug purposes
54/36 hex
converter_status (byte)
4
Refer to
actual_errors field of Error list
(section 7.2)
58/3A hex
system_state (byte)
1
Instrument state
1 - Stop
2 - Startup
3 - Measurement
5 - Standby
6 - Calibration
59/3B hex
r1
(float, inversed IEEE 754)
4
Used for maintenance service
63/3F hex
r2
(float, inversed IEEE 754
4
Used by maintenance service
67/43 hex
reserved
8